Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
18 / 49
exhalation port.
If condensation appears in the tube, remove then drain the tube. Turn down the
humidification.
12.5 Using Oxygen with the Device
Oxygen may be added at the mask connection. Please observe the instructions
listed below when using oxygen with the device.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.
Turn on the device before turning on the oxygen. Turn off the oxygen before
turning off the device. Explanation of Warning: When the device is turned off, but
the oxygen flow still exists, oxygen may accumulate within the device's enclosure
and pose a fire hazard. Turning off the oxygen before turning off the device will
prevent oxygen accumulation in the device and reduce the risk of fire. This warning
applies to CPAP devices.
Oxygen supports combustion. Keep the device and the oxygen container away
from heat, open flames, any oily substances, or other sources of ignition. DO NOT
smoke in the area near Luna
®
G3 APAP / Luna
®
G3 CPAP or the oxygen container.
• Sources of oxygen should be located more than 1 m from the device.
When using oxygen with this system, a Pressure Valve must be placed in-line with
the patient circuit between the device and the oxygen source. The pressure valve
helps prevent the backflow of oxygen from the patient circuit into the device when
the unit is off. Failure to use the pressure valve could result in a fire hazard.
When the pressure valve is installed, the Auto On function on the device will not
work. To turn on the device either breath into your mask or select the On/Off button.
Supplemental oxygen must not be used while smoking or in the presence of an
open flame.
• When using the device with an oxygen supply, check the following:
Starting therapy - ensure the device is on and blowing air before the oxygen supply
is turned on.
Stopping therapy - ensure the oxygen supply is turned off first, then the device.
This will ensure oxygen does not accumulate within the device and create a risk of
fire.
• Do not connect the device to an unregulated or high pressure oxygen source. The
pressure of oxygen source does not exceed the work pressure of the device.
12.6 Inserting the SD Card
Insert the SD card into the SD Card Slot, as shown in Fig. 12-8.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC REACTHEALTH Luna G3 APAP and is the answer not in the manual?

BMC REACTHEALTH Luna G3 APAP Specifications

General IconGeneral
BrandBMC
ModelREACTHEALTH Luna G3 APAP
CategoryRespiratory Product
LanguageEnglish

Related product manuals