Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual
35 / 49
Fig. 18-4
Fig. 18-5
(3) Hold both the cuff of the air tubing and the swivel of the mask, then gently pull
apart as shown in Fig. 18-6.
Fig. 18-6
(4) Soak the tubing in 2% soapy water at 35°C to 40°C for 10 minutes, then clean the
inside and outside surfaces with a soft brush 5 times, and rinse it with sterile water for
5 minutes.
(5) Gently tap the tubing to remove excess moisture from the connector ports. Hang
the tubing so that both ends of the opening face the floor, let the tubing naturally
dry, out of direct sunlight.
(6) Inspection
Carry out a visual inspection of the components. If there are any obvious signs of
deterioration (holes, tears or cracks, etc.), these parts should be discarded and
replaced. A slight fade may occur, which is acceptable.
WARNINGS!
• Please clean the tubing by hand.
•The tubing should be cleaned daily.
• If the heated tubing is damaged (such as broken hole, tear, exposed heating wire)