RESmart GII BPAP System Y / U Series User Manual V2.1
27 / 42
setting is “20 hPa”.
Off, 5 ~ 26 hPa applies to Y-25T and
U-25T, in 0.5 hPa increments, the default
setting is “25 hPa”.
Off, 5 ~ 31 hPa applies to Y-30T, U-30T
and U-30AT, in 0.5 hPa increments, the
default setting is “30 hPa”
When the airflow is on, an audible alert
will sound if the airway pressure is below
the warning limen; the screen will display
“Low Pressure!!”.
Note: The limens for different models:
Off,3 ~ 19 hPa applies to Y-20T and
U-20T, in 0.5 hPa increments, the default
setting is “4 hPa”.
Off, 3 ~ 24 hPa applies to Y-25T and
U-25T, in 0.5 hPa increments, the default
setting is “4 hPa”.
Off, 3 ~ 29 hPa applies to Y-30T, U-30T
and U-30AT, in 0.5 hPa increments, the
default setting is “4 hPa”
When the airflow is on, an audible alert
will sound if the respiratory rate is below
the limen; the screen will display “Low
RR!!!”.
Setting range: Off, 4 ~ 40 BPM, in 1
BPM increments, the default setting is “6
BPM”.
Note: This function is available under the
work mode of S/T or T
When SpO
2
Kit is applied, an audible alert
will sound when the value of SpO
2
is
lower than the warning threshold; the
screen will display “Low SpO
2
!!!”.
Setting range: Off, 70% ~ 100%, in
1% increments, the default setting is
“85%”.
Note: This function is available only
when the device is equipped with SpO
2
Kit
When the airflow is on, an audible alert
will sound if the air leak rate exceeds 150
L/min; the screen will display “Leak!!”.
The alarming duration time is no less than
30s.