RESmart GII BPAP System Y / U Series User Manual V2.1
28 / 42
When the airflow is on, an audible alert
will sound if the vents of the mask are
blocked; the screen will display “Mask
Blocked!!”
When the airflow is on, an audible alert
will sound if the minute volume is below
the warning limen; the screen will display
“Low MV!!”.
Setting range: Off, 1 ~ 30 L/min, in 1
L/min increments, the default setting is
“1 L/min”
If the voltage supplied by power adaptor
is lower than 22V, an audible alert will
sound and the screen will display “Low
Input Voltage!!”
When the airflow is on, an audible alert
will sound if the respiratory rate exceeds
the threshold; the screen will display
“High RR!!”.
Setting range: Off, the setting value of
Low RR ~ 80 BPM, in 1 BPM increments,
the default setting is “40 BPM”.
Note: This function is avaliable under the
work mode of S/T or T
When humidifier is applied, an audible
alert will sound when the humidifier fails
to work in 10 minutes; the screen will
display “Humidifier Failure!!”
When the Filter Alert feature is enabled,
an audible alert will sound if the preset
replacement time reaches but without
replacing the air filter; the screen will
display “Please Change Filter!”. The
default setting is “Off”.
The screen will display “SD Card Full!” if
the SD card has reached its maximum
capacity
The screen will display “Reinsert SD
card!” if the SD card fails to work
Note: the delay time of alarming system of this device is no more than 1s.