CHANGE CARE UNIT
1. From the Home screen, scroll to “Change Care Unit”
using and select with .
NOTE: Care Unit may be changed while the pump is
infusing or stopped.
2. Scroll to desired Care Unit using
and select
with .
3. Pump will display conrmation message that Care
Unit has been changed. Press to conrm.
4. If current values exceed soft limit, editor will be
displayed. Respond to soft limit message.
NOTE: Pump will display message if Care Unit was not
changed and reason why Care Unit cannot be changed
(i.e. drug or concentration not available, programming
exceeds hard limits).
BASIC INFUSION
1. Scroll to “Basic infusion” on Care Unit, Patient Pro-
le, Drug Category or Drug selection screens. “Basic
infusion” may also be accessed in the SECondary
menu.
2. The pump requires responding to two prompts that
exiting the drug library has no safety limits.
3. Enter two of the following three parameters using
: VTBI, Time and Rate. Press to conrm
values. VTBI is required to start the infusion.
4. Press
to begin infusion.