Aesculap
®
Acculan 4 / ELAN 4 electro
Aesculap Power Systems
en Instructions for use/Technical description
Attachments for small drill GA344 and drill GA844
USA Note for U.S. users
This Instructions for Use is NOT intended for United States users. Please
discard. The Instructions for Use for United States users can be obtained
by visiting our website at www.aesculapusa.com. If you wish to obtain a
paper copy of the Instructions for Use, you may request one by contacting
your local Aesculap representative or Aesculap's customer service at 1-
800-282-9000. A paper copy will be provided to you upon request at no
additional cost.
de Gebrauchsanweisung/Technische Beschreibung
Aufsätze für kleine Bohrmaschine GA344 und Bohrmaschine GA844
fr Mode d’emploi/Description technique
Embouts pour petite perceuse GA344 et perceuse GA844
es Instrucciones de manejo/Descripción técnica
Cabezales para taladro pequeño GA344 y taladro GA844
it Istruzioni per l’uso/Descrizione tecnica
Terminali per la perforatrice piccola GA344 e la perforatrice GA844
pt Instruções de utilização/Descrição técnica
Cabeçotes para máquina de furar pequena GA344 e máquina de furar
GA844
nl Gebruiksaanwijzing/Technische beschrijving
Opzetstukken voor kleine boormachine GA344 en boormachine GA844
sv Bruksanvisning/Teknisk beskrivning
Tillsatser för liten borrmaskin GA344 och borrmaskin GA844
fi yttöohje/Tekninen kuvaus
Istukat pienelle porakoneelle GA344 ja porakoneelle GA844
et Kasutusjuhend/Tehniline kirjeldus
Otsakud väikese puurimismasina GA344 ja puurimismasina GA844 jaoks
ru Инструкция по примению/Техническое описание
Насадки для малой дрели GA344 и дрели GA844
cs Návod k použití/Technický popis
Násadce pro malou vrtačku GA344 a vrtačku GA844
pl Instrukcja użytkowania/Opis techniczny
Nasadki do małej wiertarki GA344 i wiertarki GA844
sk Návod na použitie/Technický opis
Nadstavce na malú vŕtačku GA344 a vŕtačku GA844
tr Kullanım Kılavuzu/Teknik açiklama
Küçük matkap GA344 ve küçük matkap GA844 için başlıklar

Table of Contents

Other manuals for Braun Aesculap Acculan 4

Questions and Answers

  • S
    Samuel GarciaSep 12, 2025
    What to do if the tool on my Braun Aesculap Acculan 4 won't attach or detach?
    • T
      Tina PerezSep 12, 2025
      If the tool on your Braun Medical Equipment does not attach or detach, ensure you are using a tool that is compatible with the attachment. If the tool connector is deformed or defective, use a new tool. If the coupling on the attachment is deformed or defective, have the manufacturer repair the attachment. A dirty tool coupling can also cause this issue; try cleaning the tool or using a new one, and also clean the attachment.
  • M
    Mark HughesAug 25, 2025
    What to do if Braun Aesculap Acculan 4 Medical Equipment product becomes too hot?
    • T
      tbraunAug 25, 2025
      If your Braun Medical Equipment is overheating, it might be due to excessive use; ensure you are following the operating instructions. Incorrect reprocessing or maintenance could also be the cause, so adhere to the preparation and care guidelines, and lubricate the product before sterilization. Other potential causes include fall damage or a blunt tool. If the product has fall damage, have it repaired by the manufacturer. If a tool is blunt, replace it.
  • S
    Steven ThomasAug 25, 2025
    Why Braun Aesculap Acculan 4 has insufficient power?
    • R
      Ryan Brown PhDAug 26, 2025
      Insufficient power in your Braun Medical Equipment could stem from a defective product, in which case you should follow the operating instructions for preparation and care, lubricate the product before sterilization, and have the manufacturer repair it. It could also be due to a large amount of heat generated after a short time, so follow the operating mode instructions and have the manufacturer repair the product. A blunt tool can also cause this issue; if so, change the tool.
  • A
    Arthur Mccall DDSSep 23, 2025
    What to do if the speed control knob is not working on my Braun Medical Equipment?
    • R
      Rachel YoungSep 23, 2025
      If the speed control knob on your Braun Medical Equipment is not working, first ensure that the product is not in the locked OFF position; if it is, turn the safety catch to the ON position. If the knob sticks or is defective, have the manufacturer repair the product.
  • E
    Evelyn WebsterSep 12, 2025
    Why is the Kirschner wire not turning in my Braun Aesculap Acculan 4 Medical Equipment?
    • E
      Eric GibsonSep 12, 2025
      If the Kirschner wire is not turning in your Braun Medical Equipment, first ensure that the Kirschner wire chuck is set correctly by adjusting the sleeve to the correct Kirschner wire diameter. Also, check that the clamping lever is closed. If it is open, close it. If the clamping lever was not pulled back with sufficient force, pull it back forcefully and hold it.
  • D
    Dustin LinSep 16, 2025
    What to do if Kirschner wire is not turning in my Braun GB641R Medical Equipment?
    • J
      Jason GomezSep 16, 2025
      If the Kirschner wire is not turning in your Braun GB641R Medical Equipment, first, ensure that the Kirschner wire chuck is set correctly by adjusting the sleeve to the right Kirschner wire diameter. Also, make sure the clamping lever is closed. If it is open, close it. Finally, ensure that you are pulling back the clamping lever forcefully and holding it.
  • K
    Kevin AllenSep 5, 2025
    What to do if Braun Aesculap Acculan 4 saw blade has insufficient cutting performance?
    • G
      George SmithSep 5, 2025
      If you're experiencing insufficient cutting performance with the saw blade of your Braun Medical Equipment, consider the following: * **Worn saw blade teeth:** Replace the saw blade. * **Weak battery:** Charge the battery. * **Worn or defective battery:** Replace the battery. * **Overheated saw blade:** Move the saw back and forth during sawing and ensure sufficient irrigation of the saw blade.
  • M
    Misty StevensSep 5, 2025
    Why does my Braun Aesculap Acculan 4 Medical Equipment have insufficient power?
    • W
      William RiveraSep 6, 2025
      If the Braun Medical Equipment has insufficient power, here are a few possible reasons: * The product may be defective. Follow the operating instructions for preparation and care, lubricate the product before every sterilization, and have the manufacturer repair the product. * The product is operating in left rotation mode. If using a serrated tool, operate in right rotation. * A defective attachment is being used. Have the manufacturer repair the attachment. * The tool is blunt. Change the tool.
  • T
    Tracy MedinaSep 18, 2025
    How to fix oil spray adapter if it is not attachable to my Braun Aesculap Acculan 4?
    • V
      Vanessa FisherSep 18, 2025
      If the oil spray adapter is not attachable to your Braun Medical Equipment, ensure you are using the compatible oil spray adapter for GA344.
  • Z
    Zachary MorenoAug 30, 2025
    Why Braun Aesculap Acculan 4 Medical Equipment dermatome blade has insufficient cutting performance?
    • J
      Jessica SmithAug 30, 2025
      If the dermatome blade on your Braun Medical Equipment has insufficient cutting performance, the dermatome blade may be blunted, so replace it. Also, the dermatome may need to be adjusted, so have the manufacturer repair the product.

Summary

Applicable To

General Information

Intended Use

Describes the primary purpose and application of the GA344 and GA844 drills with attachments.

Main Functions and Design Characteristics

Details the key features, speed ranges, and rotational direction capabilities of the system.

Operating Modes

Outlines specific operating modes, including drilling, reaming, sawing, and screwing cycles.

Safe Handling

General Safety Warnings

Provides crucial warnings regarding injury, damage, and misuse of the product.

Product Description

Scope of Supply

Lists the contents of the product package and included accessories.

Components Required for Operation

Identifies the necessary base units and tools needed for the product's function.

Operating Principle

Explains how the attachments connect to the drill and how the motor speed is managed.

Working with the Device

System Set-up

Guides the user through connecting accessories and setting up the device.

Function Checks

Ensures the proper function of the attachment and tool before and during use.

Safe Operation

Covers essential safety precautions during the operation of the device and its accessories.

Validated Reprocessing Procedure

General Safety Notes for Reprocessing

Adheres to regulations and guidelines for safe and effective reprocessing of medical devices.

General Information on Reprocessing

Provides information on residue removal, chemical compatibility, and material care during cleaning.

Preparations at the Place of Use

Steps to take before cleaning, including detaching components and removing residues.

Preparation Before Cleaning

Instructions for preparing the product and accessories for automatic cleaning/disinfection.

Cleaning and Disinfection

Details the cleaning and disinfection methods, including specific safety instructions.

Manual Cleaning with Wipe Disinfection

Mechanical Cleaning;Disinfection with Manual Pre-cleaning

Manual Pre-cleaning with a Brush

Step-by-step guide for manual pre-cleaning using a brush and running water.

Mechanical Alkaline Cleaning and Thermal Disinfection

Details the process for machine-based cleaning and disinfection using alkaline solutions.

Inspection, Maintenance, and Checks

Post-Reprocessing Checks

Instructions for checking cleanliness, damage, function, and operation after cleaning.

Packaging

Steam Sterilization

Storage

Maintenance

Troubleshooting

Technical Service

Accessories and Spare Parts

Technical Data

Classification (Directive 93;42;EEC)

Details the product's classification according to European medical device directives.

Performance Data and Standards

Provides information on product performance, testing cycles, and applicable standards.

Operating Mode

Outlines specific operating modes, including drilling, reaming, sawing, and screwing cycles.

Environmental Conditions

Specifies the required temperature, humidity, and pressure conditions for operation and storage.

Disposal

Braun Aesculap Acculan 4 Specifications

General IconGeneral
BrandBraun
ModelAesculap Acculan 4
CategoryMedical Equipment