► For Kirschner wire attachment GB894R: Set the adjustment sleeve to
the greatest Kirschner wire diameter.
► Insert the product in the correct position into the ECCOS fixation, see
Fig. E.
► Then connect the interior rinse device to the ECCOS holder and connect
to the rinse connector of the cleaning/disinfection device/rinse cart.
► After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.8 Inspection, maintenance and checks
► Allow the product to cool down to room temperature.
► Spray product after every cleaning and disinfection with oil spray
adapter 23 GB600880 (green) approx. 2 sec with Aesculap STERILIT
Power Systems oil spray GB600, see Fig. F.
Note
Aesculap additionally recommends occasional spraying of moving parts
(such as knobs, couplings, cover lids) with the Aesculap STERILIT-Power-
Systems oil spray.
► Check the product after each cleaning and disinfection for the follow-
ing: cleanliness, damage, function, abnormal operation noise, excessive
heat or heavy vibration.
► Set aside the product if it is damaged.
7.9 Packaging
► Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for Aesculap-
ECCOS holder system).
► Insert the products in the correct position into the ECCOS fixations,
see Fig. E.
► Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
► Ensure that the packaging will prevent a recontamination of the prod-
uct.
7.10 Steam sterilization
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
► Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
► Use a validated sterilization method:
– Steam sterilization using fractional vacuum process
– Steam sterilizer DIN EN 285 and validated pursuant to
DINENISO17665
– Sterilization using the fractionated vacuum process at 134 °C,
holding time 5 min
When sterilizing multiple products in one steam sterilizer:
► Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.
7.11 Storage
► Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8. Maintenance
To ensure reliable operation, the product must be maintained at least once
a year.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9. Troubleshooting list
► Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue Cause Detection Remedy
Product becomes
too hot
Excessive use Attachment very hot Follow operating instructions (operating mode).
Defective attachment
drive/ball bearings
Attachment very hot Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steriliza-
tion.
Have the manufacturer repair the attachment.
Fall damage, product
defective
Attachment very hot Have the attachment repaired by the manufacturer.
Blunt tool Heating of tool and product Change tool.
Insufficient
power
Product is operating in
left rotation mode
Serrated tool run in counterclockwise
mode
Use serrated tools only in clockwise mode.
Attachment defective Attachment very hot Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every steriliza-
tion.
Follow operating instructions (operating mode).
Have the attachment repaired by the manufacturer.
Blunt tool Tool blades worn out Change tool.