Guidance and manufacturer’s declaration electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for it’s internal func-
tion. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions
IEC 61000-3-2
Not Applicable
The ME equipment is
solely battery powered.
Voltage fluctuations /
flicker emissions
Not Applicable
Guidance and manufacturer’s declaration electromagnetic immunity
The ME is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level
Electromagnetic environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kwww Air
Complies
Floors should be wood, con-
crete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the
shielded location from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance calculation
provided below. If a known
transmitter is present the
specific distance can be
calculated using the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not Applicable
(no electrical
cabling)
Power frequency magnetic
field
IEC 61000-4-8
3 A/m Complies
Power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Voltage dips, short
interrupts and voltage
variations on power
supply input lines
IEC 61000-4-11
>95% dip 0.5
cycle
60% dip 5 cycles
70% dip 25
cycles
95% dip 5 sec.
Not Applicable
The ME equipment
is solely battery
powered.
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
80 MHz to
800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment
is solely battery powered.
Surge
IEC 61000-4-5
±1kV
differential
±2kV common
Not Applicable
d=[ ]
V1
3.5
P
d=[ ]
E1
3.5
P
d=[ ]
E1
7
P
d=[ ]
V1
3.5
P
d=[ ]
E1
3.5
P
d=[ ]
E1
7
P
d=[ ]
V1
3.5
P
d=[ ]
E1
3.5
P
d=[ ]
E
1
7
P

Other manuals for Braun NTF 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun NTF 3000 and is the answer not in the manual?

Braun NTF 3000 Specifications

General IconGeneral
BrandBraun
ModelNTF 3000
CategoryThermometer
LanguageEnglish

Related product manuals