Guidance and manufacturer’s declaration – electromagnetic immunity
The ME is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete,
or ceramic tile. If floors
are covered
with synthetic material, the relative
humidity should be at least 30%
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the
shielded location from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Separation distance calculation
provided below. If a known
transmitter is present the specific
distance can be calculated using
the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not applicable
(no electrical
cabling)
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not applicable
The ME equipment is solely
battery powered.
Surge
IEC 61000-4-5
±1kV differential
±2kV common
Not applicable
Remove the battery from the instrument if it is not required for extended periods of time in
order to avoid damage to the thermometer resulting from a leaking battery.
Please do not dispose of the product in the household waste at the end of its
useful life.
To protect the environment, dispose of empty batteries at appropriate collection
sites according to national or local regulations.