EasyManuals Logo

Braun NTF 3000 User Manual

Braun NTF 3000
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
English
17
Guidance and manufacturer’s declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for it’s internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions
IEC 61000-3-2
Not applicable
The ME equipment is solely battery powered.
Voltage fluctuations /
flicker emissions
Not applicable
Non-life support equipment separation distance calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
80 MHz to
800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
d=[ ]
V1
3.5
P
d=[ ]
E1
3.5
P
d=[ ]
E1
7
P
Power frequency magnetic
field
IEC 61000-4-8
3 A/m Complies
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Voltage dips, short
interrupts and voltage
variations on power
supply input lines
IEC 61000-4-11
>95% dip 0.5
cycle
60% dip 5 cycles
30% dip 25
cycles
95% dip 5 sec.
Not applicable
The ME equipment is solely battery
powered.

Other manuals for Braun NTF 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun NTF 3000 and is the answer not in the manual?

Braun NTF 3000 Specifications

General IconGeneral
Measurement Time2 seconds
DisplayLCD
Silent ModeYes
Fever GuidanceYes, color-coded display
Age Precision TechnologyYes
Auto Power OffYes
WaterproofNo
TypeForehead thermometer
Battery TypeAAA (2x)
Measurement Range34.4°C - 42.2°C (93.9°F - 108.0°F)
Accuracy±0.2°C (±0.4°F)