7
English
Note: With a preferred age range, an alternate age range can be selected during each use, by
pressing the Age Precision
Technology button when the age range icon appears, after the the
thermometer is turned ON.
F. To remove the preferred age range, open the battery door, remove the batteries, wait 30
seconds and re-insert the batteries.
IMPORTANT: To ensure accurate readings, you must always take a temperature reading in the age
range that corresponds to the individual’s age.
4
Position correctly with PositionCheck
Position thermometer on or up to 2.5 cm (1”) away from
the center of the forehead, just between the eyebrows.
When the thermometer is ready and on or 2.5 cm away
from the forehead. A horizontal line of dashes (“---”) will
appear on the screen.
For non-contact readings, the yellow aiming light will
show you where you are aiming.
If the eyebrow area is covered with hair, sweat or dirt,
please clean the area beforehand to improve the
reading accuracy. It is important to hold the
thermometer and the forehead steady during
measurement. Movement will impact the temperature
reading.
5
Take temperature
Three dashed lines mean the thermometer is ready to
take a temperature. Press the temperature button. The
display will show a dashed line animation while it takes
a reading.Once the reading is complete a conrmation
beep will be heard and the appropriate, fever light color
will be displayed on screen.
If the device is positioned too far away from the forehead, it will prompt you to
move closer by displaying the letters “0-2 cm. Slowly move the device toward
the forehead until the dashed line animation starts and a reading is displayed.
0 - 2.5 cm Away
Yellow light
will show
you where
you are
aiming.