The purpose of this document is to provide the necessary information to check the accuracy of the PRO 6000
ear thermometer. The document will be used to standardize the procedure of checking the thermometer accuracy.
2.0 General information
The Braun ThermoScan® thermometer is a clinical grade ear thermometer indicated for the intermittent measurement
of human body temperature in patients of all ages in a professional use environment.
2.1 Technicalspecications
Displayed temperature range: 20-42.2°C (68-108°F)
Operating ambient temperature range: 10-40°C (50-104°F)
Display resolution: 0.1°C or °F
Accuracy for displayed temperature range:
With probe cover:
±0.2°C (±0.4°F) (35.0-42°C) (95-107.6°F)
±0.3°C (±0.5°F) (outside this temperature range)
Without probe cover:
±0.1°C (35.0-42°C)
±0.2°C (outside this temperature range)
Clinical repeatability: <0.3°C (0.57°F)
Warm up time: initial start-up time: 3-4 seconds
Measurement time: 2-3 seconds
Automatic power down 10 seconds
Battery life 6 months/1000 measurements
Battery type 2*MN 1500 or 1.5V AA (LR6)
Thermometer dimension 6”x1.7”x1.3” (152mm x 44mm x 33mm)
Thermometer weight 3.6 oz (100g) without batteries
Long term storage ranges
Storage temperature -25-55°C (-13-131°F)
Storage humidity 15-95% RH non-condensing
Shock withstands drops of 1 meter (3 feet)
Pressure 700-1060hPA (0.7-1.06 atm)
Thethermometerisspeciedtooperateat0.7-1.06atmosphericpressure.