4
5.0 Tobegincalibrationvericationofthethermometer
Caution: Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to allow for thermal accommodation.
Caution: The ambient temperature must be stable and within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F)
Caution: Setup the 9600 Plus away from sunlight, drafts, and other sources of heat or cold. Fluctuations will
impact the calibration check.
Check the probe tip lens window for displacement. There shouldn’t be any gap in the
seam formed between bezel and the lens window. If a gap exists the unit is damaged
and must be replaced.
Gently wipe the probe tip with a cotton swab slightly moistened with Isopropyl or Ethyl
alcohol using care not to displace the lens window; remove excess alcohol with a clean
cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to
clean the probe window.
Place the Braun ThermoScan PRO 6000 in calibration mode using the following steps:
a. With the thermometer in sleep mode, press and release the C/F button
b. Immediately (while all segments of the LCD are illuminated during self-test) press
and hold the C/F button and the Mem button.
c. After 5 seconds the thermometer will begin a long beep. During the beep release the
Mem button and the C/F button. The thermometer is now in Calibration check mode,
indicated by “CAL” being shown on the screen for 0.25 seconds every second.
Afterenteringcalibrationmode(CAL),waitfor1minutebeforetakingthersttemperature.
(Note: the PRO6000 must be left on a table and not kept in the hand during this time)
Apply a new probe cover. If applied properly a short beep will sound
andtheashingprobecovericonwilldisappearfromthedisplay.
Mem
C / F
Mem
C / F
Mem
C / F
1 min