4
5.0 Tobegincalibrationvericationofthethermometer
Caution: Store thermometers for testing in the same room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to allow for thermal accommodation.
Caution: The ambient temperature must be stable and within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F)
Caution: Setup the 9600 Plus away from sunlight, drafts, and other sources of heat or cold. Fluctuations will
impact the calibration check.

Check the probe tip lens window for displacement. There shouldn’t be any gap in the
seam formed between bezel and the lens window. If a gap exists the unit is damaged
and must be replaced.

Gently wipe the probe tip with a cotton swab slightly moistened with Isopropyl or Ethyl
alcohol using care not to displace the lens window; remove excess alcohol with a clean
cotton swab, and let air dry for 5 minutes. Do not use any chemical other than alcohol to
clean the probe window.

Place the Braun ThermoScan PRO 6000 in calibration mode using the following steps:
a. With the thermometer in sleep mode, press and release the C/F button
b. Immediately (while all segments of the LCD are illuminated during self-test) press
and hold the C/F button and the Mem button.
c. After 5 seconds the thermometer will begin a long beep. During the beep release the
Mem button and the C/F button. The thermometer is now in Calibration check mode,
indicated by “CAL” being shown on the screen for 0.25 seconds every second.

Afterenteringcalibrationmode(CAL),waitfor1minutebeforetakingthersttemperature.
(Note: the PRO6000 must be left on a table and not kept in the hand during this time)

Apply a new probe cover. If applied properly a short beep will sound
andtheashingprobecovericonwilldisappearfromthedisplay.
3.000SCALE
Mem
C / F
Mem
C / F
Mem
C / F
1 min

Other manuals for Braun ThermoScan PRO 6000