A. Factory Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 27 / 96
Instrument Configuration D8 ENDEAVOR
Instrument Details Set Values for Validation
Diffractometer D8 ENDEAVOR
1)
Tube Power: 40kV / 40 mA (for FL tubes)
2)
Tube Power: 40kV / 30 mA (for FF tubes)
2)
Vertical
goniometer
Theta/Theta
Height:
150
Radius:
200,5
Rotating Sample position A1
Monochromati-
sation
Kβ-Filter
Discriminator Settings
(LYNXEYEXE/ LYNXEYEXE‑T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr. [V] Veto Thr. [V]
Slits Fixed divergence Divergence slit: 0.3° (approx. 0.6mm)
Variable divergence Divergence slit: 0.3° (approx. 0.6mm)
Primary / secondary axial Soller
slit
Primary and secondary axial Soller slit: 2.5°
Air Scatter Screen Set mode to Automatic for motorized air
scatter screen. Position the air scatter screen
approx. 4mm above the sample surface for
manual air scatter screen or use adjustment
tool
LYNXEYE XE
LYNXEYE XE-T
LYNXEYE‑2
SSD160‑2
LYNXEYEXE User Manual DOC-M88-EXX240
LYNXEYEXE‑T User ManualDOC-M88-EXX239
LYNXEYE‑2 User Manual DOC-M88-EXX305
SSD160-2 User Manual DOC-M88-EXX306
1)
including D8ENDEAVORECO
2)
40kV / 25mA in case of D8 ENDEAVORECO

Table of Contents

Other manuals for Bruker D8 Discover

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker D8 Discover and is the answer not in the manual?

Bruker D8 Discover Specifications

General IconGeneral
BrandBruker
ModelD8 Discover
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals