C. Factory Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 43 / 96
Part I: Instrument Details
Instrument Configuration D8 ADVANCE
Instrument Details Set Values for Validation
Diffractometer D8 ADVANCE
1)
Tube Power: 40kV / 40 mA (for FL tubes)
2)
Tube Power: 40kV / 30 mA (for FF tubes)
2)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
Radius:
Primary radius
Secondary radius
....................[mm]
....................[mm]
Sample stage Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Compact UMC
Monochroma-
tisation
Primary Göbel mirror 60 mm Exit slit 1.2mm
Primary TWIN / TRIO parallel
beam
Exit slit 1mm
Discriminator Settings
(LYNXEYE XE/ LYNXEYE
XE-T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr.
[V]
Veto Thr. [V]
Slits Primary axial Soller slit 2.5°
Secondary equatorial Soller
slit
0.2°
Detector Scintillation counter
LYNXEYE, SSD160
LYNXEYE XE
LYNXEYEXE‑T
LYNXEYE‑2
SSD160-2
LYNXEYE User Manual DOC-M88-EXX095
LYNXEYEXE User Manual DOC-M88-EXX240
LYNXEYEXE‑T User ManualDOC-M88-EXX239
LYNXEYE‑2 User Manual DOC-M88-EXX305
SSD160-2 User Manual DOC-M88-EXX306
Solid state detector
(SOL‑XE)
SOL‑XE User Manual DOC-M88-EXX113

Table of Contents

Other manuals for Bruker D8 Discover

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker D8 Discover and is the answer not in the manual?

Bruker D8 Discover Specifications

General IconGeneral
BrandBruker
ModelD8 Discover
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals