3 Technical data
This section introduces the reader to the Recirculating Chiller and its specifications. It contains the
technical data, requirements and performance data.
3.1 Technical data
Technical data
F-100 F-105
Dimensions (W x H x D) 280 x 500 x 420 mm 280 x 500 x 420 mm
Weight 28 kg 30 kg
Power consumption (max.) 850 W 850 W
Supply voltage 230 VAC ±10%
115 VAC ± 10 %
230 VAC ±10%
115 VAC ± 10 %
Fuse
230 V
115 V
6.3 AT
8 AT
6.3 AT
8 AT
Frequency 50/60 Hz 50/60 Hz
Environmental conditions For indoor use only
- Temperature 5–35°C 5–35°C
- Altitude up to 2000 m
- Humidity Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50% relative humidity at 40°C
Temperature display No display digital, resolution 0.1°C
Overvoltage category II II
Degree of protection IP20 IP20
Pollution degree 2 2
Cooling capacity at 15°C --- 530 W
Cooling capacity at 10°C 300 W 390 W
Cooling capacity at 0°C --- 120 W
Cooling capacity at -10°C --- 10 W
Refrigerant R134a R134a
Set temperature range x +10°C –10°C … +25°C
Temperature regulation accuracy ±2°C ±1°C
Tank volume 3 L 3 L
Tubing connection 8 mm (GL14) 8 mm (GL14)
Pump ow rate 2.5 L/min 2.5 L/min
Pump pressure (max.) 0.6 bar 0.6 bar