6 Operation
20
F-100 / F-105 Operation Manual, Version C
Rear side
1
8
2
7
5
6
3
4
1 Cooling medium ow in
2 Cooling medium ow out
3 Power socket
4 Communication cable socket, RS485
(not available for F-100 models)
5 Fuse
6 Slots for air ow out
7 Antistatic wheels (not lockable)
8 Drainvalve, to empty the cooling medium tank
6.2 Preparing for use
Prerequisites
All parts must be clean and free of damage.
Close the drainvalve.
6.2.1 Installing the hoses
!
WARNING
Stumbling or falling due to improper installation of the cables and hoses.
The length of the cables and hoses should be kept as short as possible
If possible, avoid installing cables and hoses in aisle areas
If installation of cables and hoses in aisle areas is unavoidable, use an adequate protective
pad in order to avoid the danger of stumbling and damage
Connect out flow and return flow connections of the Recirculating Chiller with the condenser of the
rotary evaporator.
NOTE
Only use hoses which can withstand min. –10°C and 2bar pressure and secure them with the usual
hose clamps.
Connect Chiller outlet (OUT) to the second condenser installed on the pump.
Connect the outlet of the second condenser to the Rotavapor
®
Condenser.
Connect the outlet of the Rotavapor
®
Condenser to the inlet (IN) of the Recirculating Chiller.