3 Technical data
13 F-100 / F-105 / F-108 / F-114 Operation Manual, Version E
Technical data (100 V)
F-100 F-105
Power consumption (max.) 850 W 850 W
Supply voltage 100 VAC ±10% 100 VAC ±10%
Fuse 10 AT 10 AT
Frequency 50/60 Hz 50/60 Hz
Environmental conditions For indoor use only
- Temperature 5–35°C 5–35°C
- Altitude up to 2000 m
- Humidity Maximum relative humidity 80% for
temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C
Display No display digital, resolution
0.1°C
Overvoltage category II II
Degree of protection IP20 IP20
Pollution degree 2 2
Instrument dimensions
W×H×D [mm]
280×590×420 280×590×420
Weight 37 kg 40 kg
Cooling power at 15°C --- 530 W
Cooling power at 10°C 300 W 390 W
Cooling power at 0°C --- 120 W
Cooling power at -10°C --- 10 W
Refrigerant R134 R134
Temperature range x +10°C –10°C … +25°C
Accuracy ±2°C ±1°C
Tank volume 3 L 3 L
Hose connection 8.0 mm 8.0 mm
Pump ow rate 2.5 L/min 2.5 L/min
Pump pressure 0.6 bar 0.6 bar
3.2 Materials used
Materials used
Component Material designation Material code Hazardous substances
Stainless steel, powder-coated with
polyester-epoxide
Housing 1.4301
–
Copper Internal pipes and
coolers
CU –
Polyester Foil PES –
Glass-ber-reinforced epoxy resin Circuit board –
Polyvinyl chloride Cable PVC –
R134 Refrigerant 1,1,1,2-tetrauoro-
ethane
Check MSDS