6 Operation
24 F-100 / F-105 / F-108 / F-114 Operation Manual, Version E
6.5 Working with BUCHI vacuum controller
The Chiller models F-105 and above are equipped with a communication socket. Here, a BUCHI
Rotavapor setup or other appropriate laboratory instruments equipped with a BUCHI vacuum
controller (model V-850 or higher with firmware version 3.0 or higher) can be attached.
To establish communication switch on all connected devices. After bootup time the UP and DOWN
buttons of the Chillers operation panel are blocked – the Chiller can be controlled via the Vacuum
Controller buttons.
*click*
To Rotavapor /
Vacuum pump
5m
NOTE
Do not exceed max. recommended cable length (5 m). For more information about operation see
respective manual of the Vacuum Controller in use.