How to fix the value parametered for the pulse output is wrong (generated frequency is > 250 Hz) in Burkert Accessories?
- JJonathan CollinsSep 13, 2025
Increase the value of the PU parameter.
How to fix the value parametered for the pulse output is wrong (generated frequency is > 250 Hz) in Burkert Accessories?
Increase the value of the PU parameter.
What to do if the measured flow rate is above the authorized high limit in Burkert Accessories?
Check the flow rate in the pipe and its consequences on the process. If necessary, clean the flow sensor then calibrate the flow zero point.
What to do if the displayed flow rate is not nil but the flow rate in the pipe is for Burkert Accessories?
If the displayed flow rate isn't nil when the flow rate in the pipe is, verify that the flow rate in the pipe is indeed nil, check for bubbles in the pipe, inspect the chosen filter, and calibrate the flow zero point.
Why is the displayed flow rate not stable in Burkert Accessories?
If the displayed flow rate is unstable, check that there is fluid in the pipe and consider choosing a higher filter.
What if the measured flow rate is negative (although the display shows a positive flow rate) in Burkert Accessories?
Check that the reverse fluid circulation has no consequences on the process. Otherwise, mount the device on the pipe so that the arrow on the housing indicates the flow direction.
What to do if the real flow rate cannot be displayed (display is saturated) in Burkert Accessories?
Adjust the unit or the number of decimals in the UNIT parameter of the Parameters menu so that the display can show higher values.
What to do if the current value transmitted on the AO1 current output does not correspond to the displayed flow rate for Burkert Accessories?
Check the configuration of the current output and verify the OFFSET and SPAN parameters of the current output.
What to do if the DO2 and DO3 outputs do not switch whatever the displayed flow rate value in Burkert Accessories?
Check the parameters of the DO2 and DO3 relay outputs, verify the flow rate unit, and check the behavior of the outputs.
What to do if the current output transmits a value between 0 and 4 mA from Burkert Accessories?
If the current output transmits a value between 0 and 4 mA, switch the Burkert Accessories device power supply off and then on again. If the problem persists, contact the retailer.
What to do if the current output transmits 0 mA from Burkert Accessories?
If the current output transmits 0 mA, check the wiring of the current output. Also, check the fuse of the installation, replacing it if necessary, and verify the position of the sink/source switch. If the problem continues, contact the retailer.
Model | 8045 |
---|---|
Device Type | Transmitter |
Measurement Principle | Electromagnetic |
Output Signal | 4-20 mA |
Material | Stainless steel |
Lining Material Options | PTFE |
Electrode Material Options | Stainless Steel, Hastelloy |
Accuracy | ±0.5% |
Ambient Temperature Range | -20 to 60 °C |
Protection Rating | IP65 |
Certifications | ATEX |