Troubleshooting | 11
0870570604_RA0165-0305D_Ex_-0011_IM_en 31 / 36
EF FV GB IS CPL OF
Problem Possible Cause Remedy
The machine does not start. The motor is not supplied
with the correct voltage.
Check the power supply.
The motor is defective. Replace the motor.
The coupling (CPL) is de-
fective.
Replace the coupling
(CPL).
The machine does not reach
the usual pressure on the
suction connection.
Oil level too low. Top up oil.
The inlet screen (IS) is par-
tially clogged.
Clean the inlet screen (IS).
The inlet filter cartridge (op-
tional) is partially clogged.
Replace the inlet filter
cartridge.
Internal parts are worn or
damaged.
Repair the machine (con-
tact Busch).
The machine runs very nois-
ily.
Worn coupling (CPL). Replace the coupling
(CPL).
Stuck vanes. Repair the machine (con-
tact Busch).
Defective bearings. Repair the machine (con-
tact Busch).
The machine runs too hot. Insufficient cooling. Remove dust and dirt
from the machine.
Check the cooling fan.
Ambient temperature too
high.
Observe the permitted
ambient temperature.
Oil level too low. Top up oil.
The exhaust filters (EF) are
partially clogged.
Replace the exhaust filters
(EF).