17
Hold Time Settings / Run Time / Printing
II.5 HOLD TIME SETTINGS
Hold time sets the time period between when the RUN key is pressed and when
the spindle begins to rotate. This time period is normally used to ensure thermal
equilibrium of the sample and spindle. The hold time range is 0 to 999 seconds.
Press the HOLD TIME key and the current hold time will blink on the default screen.
Use the number keys to type in the required hold time and press the ENTER key.
NOTE: When the hold time is set to zero, it is not displayed on the default screen.
II.6 RUN TIME
Run time sets the time period of spindle rotation. The run time range is 0 to 999
seconds.
Press the RUN TIME key and the current run time will blink on the default screen.
Use the number keys to type in the required run time and press the ENTER key.
NOTE: Run time will be shown on the default screen only when hold time is set to
zero.
Aruntimeofzerosetstheviscometertoinniterunmode.Inthismode,thespindle
will rotate at the set speed for as long as the RUN button is pressed. When the RUN
key is released, the spindle will stop rotating.
NOTE: With a run time of zero, the hold time will not be used.
II.7 PRINTING
Pressing the PRINT key at any time sends information on test parameters to the
printer port. However, viscosity, full scale range and shear stress data will only be
printedafteritisrstdisplayedduringatestrun.
At the end of a timed speed execution, data will automatically be sent to the printer
port. To print a heading, press and hold the STOP/ESCAPE key and press the
PRINT key.
CAP 1000+ PRINT OUTPUT
VISCOSITY
(POISE)
F.S.R.
(%)
TEMP
(Deg C)
S.STRESS
(D/CM2)
S.RATE
(1/sec)
SPEED
(RPM)
TIMER
(SEC)
CONE
No.
- - 25.0 - 10000 0750 20 02
A maximum of 999 seconds can be printed when running in manual TIMER mode
(00). Over 999 seconds will print EEE.
NOTE: The TIMER (SEC) column will indicate the accumulated time of running at the
moment the print key is pressed while the cone is rotating. This time value
will not include the hold time.