16
Speed Setting / 4 Temperature Control Setting
II.3 SPEED SETTING
The CAP1000+ is a single speed viscometer. It is supplied from the factory in two
congurations:standardandcustom.Thespeedofrotationisshowninthelower
right corner of the display.
Standard
The CAP1000+ Viscometer is supplied with two speeds: 750 RPM and 900 RPM.
All CAP1000+ Viscometers are factory-set at 900 RPM. To change the setting, press
and hold the STOPkeyduringthepowerupsequence.Choose“1=SPEED(RPM)”
from the special functions screen and select the required speed. This selection will
become the default speed.
Custom
TheCAP1000+Viscometerisconguredtooperateatonespeedasspeciedatthe
time of order. This speed is indicated on the instrument serial tag.
To begin spindle rotation, press the RUN key.
II.4 TEMPERATURE CONTROL SETTING
Press the TEMP key and the current temperature setting will blink. The default
temperature on start-up is 25.0°C on low temperature models and 50°C on high
temperature models.
The temperature ranges are:
Low temperature: 5°C (or 15°C below ambient, whichever is higher) to 75°C
High temperature: 50°C to 235°C
NOTE: The temperature can be set in increments of 0.1°C.
Use the number keys to type in the required set point.
Use the ENTER key to accept the new set point.
NOTE: Thermal equilibrium of the sample and of the spindle must be considered
for best measurement results. Upon powering up the Viscometer or after
changingthetemperaturesetpoint,allowsufcienttimefortheplateto
reach the desired temperature. It is recommended to have the spindle in
contact with the plate prior to introducing the sample material to ensure that
the spindle is also at the temperature of test. BYK-Gardner recommends
using the solvent trap at all times to enhance the temperature control of the
sample material. After inserting the sample material onto the plate, lower
thespindleandsolventtrapandallowsufcienttimeforthermalequilibrium
prior to starting the test.