EasyManua.ls Logo

Cameron Barton 752 - Section 3-Installation, Startup, and Shutdown; Initial Calibration Check; Transmitter Mounting Guidelines

Cameron Barton 752
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Model 752 and 752A Differential Pressure Transmitter Section 3
Section 3—Installation, Startup, and Shutdown
Overview
This section describes the steps required to install the instrument so that it
will perform to its original factory calibration condition. Installation tasks
include
initial calibration check
mounting the transmitter
installing piping
installing eld wiring
Unpacking/Inspection
The instrument should be inspected at the time of unpacking to detect any
damage that may have occurred during shipment.
IMPORTANT: The unit was checked for accuracy at the factory. Do not change any of
the settings during examination or accuracy will be affected.
The transmitter is shipped in a polyethylene bag to protect the instrument
from contamination. Remove this bag only in a clean area.
Initial Calibration Check
The Model 752 and 752A transmitters are factory-calibrated. However, to
ensure that the calibration is intact following shipping, a calibration check
is recommended prior to operating the instrument. See Calibration, page 19,
for step-by-step instructions. Record the "as found" values and recalibrate, if
necessary.
Mounting
Mount the transmitter so that the pressure housings are in a horizontal posi-
tion and when the operator is facing the transmitter cover, the controls are on
the right side. Use mounting structures that are designed to minimize vibra-
tion and avoid resonance and/or keep resulting amplication below 33 Hz.
Support connected process tubing and conduit using the same mounting as the
instrument base to minimize relative motion of the instrument and connec-
tions.
Wall or Rack Mounting
1. Locate and drill four bracket mounting holes in the mounting surface.
2. Attach the instrument to the wall using 5/16" (8 mm) bolts, Grade 5 or
better, and torque to 10-20 ft-lb.

Table of Contents

Related product manuals