EasyManua.ls Logo

Cardioline HD+ - 9 Technical Specifications

Cardioline HD+
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HD+
9. TECHNICAL SPECIFICATIONS
33
9. TECHNICAL SPECIFICATIONS
ECG leads ........................................
12-leads (I, II, III, aVR-L-F, V1-6)
Patient cable ....................................
10 leads replaceable wire
CMRR ..............................................
115dB
DC input impedance ........................
100MΩ
A/D converter ..................................
24 bit, 32000 samples/second/channel
Sampling rate of the input stage ......
32000 samples/second/channel
Sampling rate for signal analysis ......
1000 samples/second/channel
500 samples/second/channel
Selectable via software
A/D conversion ................................
20 bit
Resolution .......................................
<1 μV/LSB
Dynamic range.................................
+/- 400 mV
Bandwidth .......................................
300 Hz (@1000 m/s)
150Hz (@500 m/s)
Pacemaker detection .......................
Hardware detection coupled with digital convolution filter
De fibrillation protection .................
AAMI/IEC standard
Front-end performance ...................
ANSI/AAMI IEC 60601-2-25:2011
Data transfer ...................................
Bluetooth 2.0+ EDR with “secure pairing”
Lead-fail detection ...........................
Independent for all leads
Compatible devices .........................
Cardioline touchecg, Cardioline cubestress.
Power supply ...................................
2 AAA standard batteries
Battery Duration ..............................
More than 500 ECGs
Dimensions ......................................
115 x 65 x 15 mm
Weight .............................................
< 90 g, including batteries
Protection against harmful ingress
of water or particular matter ...........
IP40 /IP42 with protective shell
Mechanical strength and
temperature resistance ...................
Compatible with EN 1789 (Ambulances) and EN 60601-1-11 (homecare)
requirements