Power Requirements
A power cord is provided with this equipment. All countries must use an agency-
approved power cord with a plug type suitable for the country of use. Connect the
equipment to a power source that is suitable for the voltage and current ratings at
your site.
For regions where equipment power is unstable, The use of a Uninterruptible
Power Supply (UPS) 1.5 KVA is highly recommended to ensure the correct current
supply.
The single-phase power source, with grounding, must be provided within 2.5 m
(8.0 ft) of the DRYVIEW Laser Imager.
The wire must be insulation-rated for input voltage = 600 V (ac). A dedicated line is
recommended.
Imager Power Requirements—Input Voltage
DRYVIEW 5950 Laser Imager, DRYVIEW 5700
Laser Imager
90–132 V (ac), 50/60 Hz ± 3 Hz
180–264 V (ac), 50/60 Hz ± 3 Hz
DRYVIEW DV6950 Laser Imager
90–132 V (ac), 50/60 Hz
180–264 V (ac), 50/60 Hz
Required Information Record Information from Customer Site
What is the voltage for this site?
Do we need / have an UPS on site?
AN9405 12 2024-03-06