Air Purifier Sizes 1620, 1625, 2020, 2025, 2420: Installation Instructions
Manufacturer reserves the right to change, at any time, specifications and designs without notice and without obligations.
5
A11368
Fig. 5 – Mounting Air Purifier Cabinet
A11493
Fig. 6 – Removing Filter
A11494
Fig. 7 – Removing FEM
A11545
Fig. 8 – Installation of Support Foot
5. Position the cabinet between the furnace and return air duct (Fig. 3, 
Fig. 4, and Fig. 5). A transition duct may be required.
On some furnaces, one or more screws may interfere with the ease
of removal of the purifier door. In this case, replace the interfering
screw(s) with pop rivet(s). Removing the screws without replacing
them with pop rivets may have an adverse effect on air sealing or
structural integrity of the furnace.
6. Use foam tape or silicone sealant between the furnace and the air
purifier cabinet.
7. Mounting holes are provided in the air purifier flanges for ductwork
and  furnace  attachment.  To  access  the  mounting  holes  on  the
upstream  flange  of  the  purifier  adjacent  to  the  incoming  power
wiring,  the  wiring cover  must  be temporarily removed. To do  so,
remove the  three  screws  illustrated in Fig. 9 - Step 1. Gently pull
CAUTION
!
UNIT DAMAGE HAZARD
Failure to follow this caution may result in unit damage.
Mounting  holes  are  provided  for  duct  work  and  furnace  attachment.
The screws on the down-stream side of the cabinet should be installed
so that the screw heads are inside air purifier cabinet to prevent damage
to the air purifier cartridge.