31
Err 13 (Zero Range with power on)
■ Cause
The initial zero range exceeds above 10% of maximum capacity.
■ Management
Check the load cell state.
Over
■ Cause
Too heavy object is put on platform so that it may exceed the maximum capacity.
■ Management
Never put the weight exceeding the maximum capacity.
If the load cell is damaged, replace the load cell.
Errors in Count Mode
Err 61
■ Cause
The count unit weight and sample weight are set too high.
■ Management
Set the count unit weight low or the count sample quantity high.
Err 62
■ Cause
The count unit weight and sample weight are set too low.
■ Management
Set the count unit weight high or the count sample quantity low.