EasyManua.ls Logo

CAS BW-30 - Count Mode Errors

CAS BW-30
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Err 13 (Zero Range with power on)
■ Cause
The initial zero range exceeds above 10% of maximum capacity.
■ Management
Check the load cell state.
Over
■ Cause
Too heavy object is put on platform so that it may exceed the maximum capacity.
■ Management
Never put the weight exceeding the maximum capacity.
If the load cell is damaged, replace the load cell.
Errors in Count Mode
Err 61
■ Cause
The count unit weight and sample weight are set too high.
■ Management
Set the count unit weight low or the count sample quantity high.
Err 62
■ Cause
The count unit weight and sample weight are set too low.
■ Management
Set the count unit weight high or the count sample quantity low.
CAS BW-30 - Count Mode Errors