30
ERROR MESSAGE DESCRIPTION & MANAGEMENT
Errors in Weighing Mode
Err 01
■ Cause
The Indicator can't be initialized due to platform shaking.
■ Management
Put the platform on a flat place without shaking and power on.
Err 02
■ Cause
The load cell is connected wrongly or A/D conversion part has a fault.
■ Management
Check that the load cell is well connected with main unit.
Err 03
■ Cause
The calibration isn't executed or the data of internal memory is lost due to
a electrical shock.
■ Management
Perform the required specification again in setting mode.
Err 09 (Zero Key)
■ Cause
The current weight exceeds zero range.
Err 10
■ Cause
The specified TARE weight exceeds the maximum capacity.
■ Management
Set the TARE weight lower than maximum capacity.
Or set the maximum capacity higher than the desired TARE weight in
calibration menu and perform the calibration again.