ED Service Manual
13 04/12/10
(6) Input a gravitational acceleration value by using [ZERO] key.
(7) Press [TARE] key to save the gravitational acceleration value, and “C-9 ” message
will be shown.
2.1.4. Calibration factor Setting (C-10)
(1) Pressing and holding “Calibration Switch” press [POWER] key.
(2) After “CAL” message blinks three times and shows the version of scale, it displays
“CAL-1” message.
(3) Press [ZERO] to display “C-10”.
(4) Press [TARE] key, and then “UNIT “ message and “0” will be shown. The first digit,”0”
will blink. It means calibration unit is “kg” (0 : kg, 1 : lb)
(5) Input a calibration unit by using [ZERO] key.
(6) Press [TARE] key, and then “CAPA“ message blinks.“0003” will be shown. The first
digit,”0” will blink. It means a full-capability is “3 (calibration unit, kg or lb)”
(7) Input a capability by using [ZERO] key.
(8) Press [TARE] key, and then “MCAPA“ message blinks.“0001” will be shown. The first
digit,”0” will blink. It means a mid-capability is “01 (calibration unit, kg or lb)”
(9) Input a capability by using [ZERO] key.
(10)Press [TARE] key, and then “W-dP “ message blinks.“3” will be shown. The first
digit,”3” will blink. It means a weight decimal point is “3 (will display 0.000)”
(11)Input a weight decimal point by using [ZERO] key.
(12) Press [TARE] key, and then “1d “ message blinks.“0.001” will be shown. The third
digit,”0” will blink. It means a division is “0.001 (calibration unit, kg or lb)”
(13) Input a division by using [ZERO] key.
(14) Press [TARE] key, and then “dual “ message blinks.“1” will be shown. The third
digit,”1” will blink. It means a dual interval is disable. (0 : disable, 1 : enable)”
(15) Input a dual interval enable by using [ZERO] key.
(16) Press [TARE] key to save the calibration factor, and “C-10 ” message will be shown.
2.1.5. Displaying Real A/D Value (C-5)
Display Raw AD