7. Error Messages & Solution
Error Message
on Display
Description Solution
"Err 0"
The "Err 0" occurs when scale is not stable.
Remove unstable facts.
"Err 1"
The "Err 1" occurs when a current zero
point has shifted from the last span
calibration.
Please call your CAS dealer.
"Err 2"
The "Err 2" is not a real error. Only it
prompts return CAL switch to the normal
position.
Please call your CAS dealer.
"Err 3"
The "Err 3" is an overload error. Please remove the weight.
"Err 5"
The "Err 5" means there is already tare. Remove the tare.
“Err 11
The "Err 11" means a writing error of the
internal nonvolatile memory. To recognize
this error, be sure to check the voltage on
the circuit and do calibration procedures.
If it still has "Err 11", replace the digital
module.
“Err 12”
The "Err 12" warns that the scale has lost
the parameters for weighing regulations or
has lost the factors for a digital span
calculation.
Enter each condition codes again (?).
Please try a span calibration again if still
not fixed.
“Err 14”
The "Err 14" means calibration range is not
correct.
Please call your CAS dealer.