ED Service Manual
9 04/12/10
2. Calibration
2.1. General Calibration
Pressing and holding calibration switch press [POWER] key to go to calibration mode.
User can move to other mode by using [ZERO] key in the calibration mode.
User also moves to other sub-modes for each mode by using [TARE] key.
Please simply follow below procedure to move to other mode.
(1) Calibration Mode: Pressing and holding “Calibration Switch” press [POWER] key.
(2) It displays “CAL-0” after “CAL”, and it blinks the version of scale three times.
(3) Selecting menu: press [TARE].
(4) ENTER(Setting) : [TARE] key
MODE Function
CAL 1
Display normalized AD
CAL 2
Display Keypad infomation-
CAL 3
Weight Setting Mode
“UnLoad” [TARE]
“MIDD” [TARE] after loading for 1/3 weight
“FULL” [TARE] after loading for Full weight
“MIDD” [TARE] after loading for 1/3 weight “END”
CAL 4
Option Setting ( Table 1 참조 )
CAL 5
Display filtered Raw AD
CAL 6
Function setting on each Key ( Table 2 참조 )
CAL 7
% Calibration
CAL 8
Battery calibration
CAL 9
Gravity constant
CAL 10
Set calibration factor
“Unit” Æ [TARE] Æ select 0, 1 (0:kg, 1: lb) Æ [TARE]
“CAPA” Æ [TARE] Æ select capacity Æ [TARE]
“MCAPA” Æ [TARE] Æ select mid-capacity Æ [TARE]
“W-dP” Æ [TARE] Æ Select Decimal Point Æ [TARE]
“ 1 d ” Æ [TARE] Æ Select division Æ [TARE]
“Dual” Æ [TARE] Æ Enable dual interval (0:disable, 1:enable) Æ TARE
CAL 11
Set nation(00 : OIML , 01 : NTEP , 02: KOREA)