TEST 5
FUNCTION : Key Test
KEY LCD SCREEN DESCRIPTION
HOLD KEY or
kg ◀▶ lb
KEY
: execute test 5
PUSH
ZERO
Display TEST 5 state
Pressing the key to test shows
(Power, Zero, Tare, HoLd)
4. CALIBRATION MODE
How to enter calibration mode
1. Remove the sealing bolt on the rear panel and put the switch to CAL ON.
2. Turn on the power.
3. You can see ‘CZ ST’ message.
4. Put the switch to CAL OFF and perform the sealing.
Calibration menu
CZ ST : Calibration & Set mode select
UNIT : Weight Unit Set
ZERO : Zero Calibration
MIN : Minimum Division Set
MAX : Maximum Capacity Set
BAL : Setting the Weight in Span Calibration
CAL0 ~ 3 : Span Calibration
CZ ST
Function : Calibration mode & Set mode select
KEY LCD SCREEN DESCRIPTION
ZERO :
Operate the Calibration
mode
TARE :
Operate the Set mode
CZ ST