2、Test Mode Introduction
I、 Power on to Enter Calibration Mode
Press [CAL] key, then press [ON/OFF] key and enter cal mode.
Display [ CAL ] three times, then show [C-1].Press [Zero] or Numeric keys to select
one of C1~C5 item, press [Tare] key to get into C1~C5 mode.
Notice:Because the keyboard of the scale will test the key value and the battery
voltage when enter into CAL mode, please unpress the [CAL] key, and no key will be pressed
between display [CAL] and dsplay [C1].
When keyboard can not be used, please try again.
II、 C1 Set Weighing Specification
C1:Weighing Specification Set:0~34
Display [C1 x ] Showed In Table 1 and Table2.
III、 CLEAR MEMORY
C2:EEPROM Initiation
Press [Tare] Key to Initiate EEPROM.
IV、 C3 Calibration
Display [ULoAd]
Press [Tare] Key, Confirm Zero Point。
Display [LoAd]
Load the poise, press [Tare]and wait a few seconds. If the weight loaded is just the
same as the full range, press [Tare] complete this calibration. Or you should modify the value
through Nemeric Key and Press [Tare] Key.
Notice:Calibration weighting value should not great than 10% of capacity.
V、 FUNCTION SET
C4-1:Initial Zero Range
[C4-1 0] 2% Full Range
[C4-1 1] 5% Full Range
[C4-1 2] 10% Full Range
[C4-1 3] 20% Full Range
C4-2: Zero Set Range
[C4-2 0] 2% Full Range
[C4-2 1] 5% Full Range
[C4-2 2] 10% Full Range
[C4-2 3] 20% Full Range