2

Questions and Answers

  • J
    jbarberAug 19, 2025
    What to do if my CAS PDS show Err 13?
    • K
      Katherine ClarkAug 19, 2025
      If your CAS Scales display 'Err 13', it means the zero value exceeds the initial zero range. To fix this, remove some weight from the platter and then turn the power on.
  • C
    Carrie MooreAug 21, 2025
    Why do my CAS PDS Scales show Err 08?
    • B
      Brent PonceAug 22, 2025
      If your CAS Scales display 'Err 08', it indicates that the Hi limit weight is set higher than the maximum weight. To resolve this, ensure that you set the Hi limit weight to a value that is below the maximum weight.
  • M
    Mr. David BernardAug 25, 2025
    What does Err 09 mean on CAS PDS?
    • M
      Michele WellsAug 25, 2025
      If you see 'Err 09' on your CAS Scales, it means the Low limit weight is set higher than the Hi limit weight. To correct this, set the Low limit weight to a value that is lower than the Hi limit weight.
  • A
    Anthony HudsonAug 27, 2025
    What to do if CAS Scales display Err 22?
    • M
      Melvin MoralesAug 27, 2025
      If your CAS Scales show 'Err 22', it means the weight used for span calibration is less than 10% of the maximum capacity. The weight of Span calibration should be set to over 10% of max. capacity (Check it in CAL 1)
  • M
    Megan VargasAug 28, 2025
    Why do my CAS Scales show Err 23?
    • K
      Kristina HopkinsAug 28, 2025
      If your CAS Scales display 'Err 23', it indicates that the weight for span calibration is set to over 100% of the maximum capacity. To fix this, ensure that the weight of Span calibration is set to under 100% of max. capacity (Check it in CAL 1).
  • N
    Nicholas SmithAug 30, 2025
    What to do if my CAS PDS display Err 32?
    • M
      michael60Aug 30, 2025
      If your CAS Scales display 'Err 32', it means that the number of samples is too low. To resolve this, you can either put more samples on the scale or decrease the required number of samples.
  • A
    Alison AdamsSep 1, 2025
    How to fix Err 21 on CAS PDS?
    • S
      Sharon MackSep 1, 2025
      If your CAS Scales display 'Err 21', it indicates that the resolution exceeds 1/20,000. To fix this, you should lower the resolution setting.
  • D
    Dana LewisSep 3, 2025
    What does Err 05 mean on CAS Scales?
    • B
      Barbara WatsonSep 4, 2025
      If your CAS Scales show 'Err 05', it means that the input value is outside of the acceptable range. Make sure to input a value between 32 and 255 to resolve this issue.
  • T
    Tiffany BeanSep 5, 2025
    How to resolve Err 24 on CAS Scales?
    • L
      lacostaSep 6, 2025
      If your CAS Scales display 'Err 24', it means that the value of Span is either too high or too low. To resolve this, you should adjust the resolution settings.
  • M
    Melissa HernandezSep 7, 2025
    What does Err 07 mean on CAS PDS?
    • M
      matthewskyleSep 7, 2025
      If your CAS Scales display 'Err 07', it means the Product ID is larger than 19. The range of Product ID is 0 to 19.

Summary

0. Precaution

! Attention

3. Specification

5. Front Panel

7. Test Mode

TEST 2: A;D Conversion TEST (Load Cell Test)

Performs A/D conversion test for the load cell to check its functionality and value.

TEST 3: RS - 232 TEST with Computer (RS - 232 C)

Tests the RS-232 serial communication with a computer, covering transmission and reception.

TEST 4: RS - 422;485 TEST with Computer (RS - 422;485: Option)

Tests RS-422/485 serial communication with a computer, covering transmission and reception.

TEST 5: PRINTER TEST

Tests the printer functionality by performing a printer test and checking output.

8. Function Mode

WEIGHT BACKUP (F01)

Configures settings for weight backup, including selecting options for usage and saving.

SET HOLD TYPE (F02)

Configures the hold type for weighing, allowing selection of average, maximum, or current values.

KEY LOCK (F08)

Sets up key lock functionality, allowing selection of which keys to lock and the lock state.

DEVICE Mode Menu

COM1(RS-232) USE (D02)

Configures the usage of COM1 (RS-232) port, selecting between no use, CAS, Limit, Printer, or ECR mode.

9. How to use (Weighing Mode Only)

10. Communication

How to connect PC (COM1)

Provides pinout diagrams for connecting the scale to a PC via COM1 using different port types.

11. CAS & Command Mode Protocol

CAS Protocol (22 bytes) - ASCII Code

Details the CAS Protocol structure and ASCII code representation for data transmission, including status and device information.

Limit Protocol (22 bytes)

Explains the Limit Protocol structure, covering stable/unstable states, gross/net data, and zero/limit signals.

Command Mode Protocol

Outlines the Command Mode Protocol, detailing commands for reading and writing data like tare value and device ID.

12. Error Message & Trouble Shooting

CAS PDS Specifications

General IconGeneral
BrandCAS
ModelPDS
CategoryScales
LanguageEnglish

Related product manuals