2.1.2. SPAN Calibration Setting (C-3)
Pressing and holding “Calibration Switch” press [POWER] key.
After “CAL” message blinks three times and shows the version of scale, it displays “CAL
1” message.
Press [Tare] / [Mode] to display “CAL-3”.
Press [Zero] key and then it displays “zero” message.
Press [Zero] key and then it displays “midup” message
Load middle up weight (Refer to table below) on the platform
Press [Zero] key and then it displays “FULL ” message
Load full weight on the platform
Press [Zero] key and then it displays “middn” message
Load middle down weight (Refer to table below) on the platform
Press [Zero] key and then it display “CAL 3” message
Press [Power] key to save & switch off
Table for Cal weights
Full Half weight
20 10
10 5
5 2
2 1
2.1.3. Gravity Constant Value Setting (C-9)
Current gravitational Acceleration value is set to 9.7994 m/s
2
.
Pressing and holding “Calibration Switch” press [POWER] key.
After “CAL” message blinks three times and shows the version of scale, it displays “CAL-1” message.
Press [ZERO] to display “C-9”.
Press [TARE] key, and then “ G-1“ message and “9.7994” will be shown. The first digit,”9” will blink.
Input a gravitational acceleration value by using [ZERO] key.
Press [TARE] key, and then “G-2“ message blinks.“9.7994” will be shown. The first digit,”9” will blink.
Input a gravitational acceleration value by using [ZERO] key.
Press [TARE] key to save the gravitational acceleration value, and “C-9 ” message will be shown.