INSTALLATION
Carefully follow and adhere to the instructions given in this chapter to guarantee working efficiency and
safety when the Chiller is in use.
LOCATION
Locate the Chiller in a ventilated area distant from heat sources such as radiators, air conditioning units,
deep-fryers and ovens.
Make sure the Chiller is located at a distance not less than 10 cm from the back wall to permit a good
cooling effect for the various components.
To maintain the correct internal temperature, the ambient temperature must not exceed +32°C
Adjust height and levelling using the support feet and also check the door closes. If the Chiller is not per-
fectly level working efficiency and condensation flow may be affected.
Remove the PVC protective film on both sides of the Chiller.
CONNECTING TO THE POWER SUPPLY
The T3 Chiller works off single-phase tension 230V- 50Hz.
To connect to the power supply, simply plug the power cable in the power socket or connect the Chiller
to the electrical panel, however before connecting check that:
- The socket is earthed and frequency and tension correspond to the information given on the
metal plate. If in doubt about earthing, call in a qualified electrician to check wiring.
- A suitable thermomagnetic switch must be provided between the power cable and electricity
line with at least a 3 mm break contact.
- when the Chiller is connected to the mains check that the supply tension, with the Chiller work-
ing, does not exceed or go under the rated tension by +/- 10%.
CAUTION: the manufacturer shall not be held responsible for damage or accidents caused by
negligence due to the non-observance of the recommendations and regulations given or of the local
governing standards and laws regarding electrical safety.
CAUTION: When transporting or moving the Chiller to the installation site do not push or drag it, lift
it and position it on a trolley to avoid overturning.
6