DAILY MAINTENANCE
TO BE CARRIED OUT BY INSTRUCTED SURGERY STAFF
CLEANING AND DISINFECTING YOUR SMART SUCTION UNIT
User Instructions:
1. Fill the Pulse Cleaner Bucket
with 2.5 litres of warm water
(approx temp 50°c).
4. Squeeze the bottle softly
to force the solution into the
measuring container at the top
of the bottle, until it reaches the
desired mark (10ml general
cleaning or 20ml disinfecting).
7. When all the solution has
been poured into the Pulse
Cleaner Bucket, replace the red
cap and then with a disposable
cloth, wipe away any residue
from the bottle.
2. Each line on the inside of
the Pulse Cleaner Bucket
represents 1 litre, so fill it up to
halfway between the 2nd and
3rd line.
5. If too much solution has
entered into the container it can
be poured back into the bottle
by tipping the bottle towards
the small tube on the side of the
bottle.
8. Connect your suction hoses
to the Pulse Cleaner Bucket
(adaptors may be required),
pulsed aspiration of the solution
will happen automatically.
3. Place the Puli Jet on a level
and steady surface and slightly
unscrew the red cap, but do not
completely unscrew it.
6. Unscrew and remove the red
cap fully and pour the solution
into the Pulse Cleaner Bucket
making sure that no extra
solution travels up the pipe
from the main bottle.
9. Once pulsing has stopped
please remove the suction
hoses from the Pulse Cleaner
Bucket.
Please note: there may be a small
amount of solution left at the bottom
of the Pulse Cleaner Bucket, this
can be poured down the spittoon or
aspirated manually.
It is recommended that you
continue to aspirate air for several
minutes before turning off the
suction system. This ensures that
the disinfectant reaches the suction
unit in the plant room.