DAILY MAINTENANCE
TO BE CARRIED OUT BY INSTRUCTED SURGERY STAFF
CHANGING THE AMALGAM CONTAINER
MICRO SMART / MICRO SMART CUBE
1. Turn off the unit with the switch on the control panel.
2. Turn the handle underneath the amalgam container
clock-wise until the container reaches the bottom of the
thread.
3. Gently pull the container outward.
4. Top up the amalgam container to within 5mm of the rim
with neat Puli Jet Plus. Place the container on a solid
surface and carefully place a new lid on top, twist and lock
in position. Then push in the small plastic bung and twist
to lock. The container is now airtight and safe for disposal*.
If re-using the original amalgam container, empty the
contents into an approved amalgam waste container**
5. Before replacing or refitting the new container treat the
o-ring seal with a silicon based lubricant (not vaseline).
6. Replace with a new container (043132), relocate the new
container into the stainless steel guide making sure the
plastic disc fits within the recess at the base of the
amalgam pot.
7. Push the container forward and turn the handle anti
clock-wise until container is fully located and sealed at
the top of the lever.
1
2
6
7