DAILY MAINTENANCE
TO BE CARRIED OUT BY INSTRUCTED SURGERY STAFF
CHANGING THE AMALGAM CONTAINER
TURBO SMART / TURBO SMART CUBE
1. Turn off the unit with the switch on the control panel
2. Turn the vent valve on the top of the amalgam separator
anti clock-wise to open it so you can empty the excess
liquid from the amalgam container. Wait 15 seconds.
Turn the drain tap (left hand side of amalgam container) to
the vertical position allowing the liquids to drain into the pot.
3. Gently hold the collection container and slowly lift the
stainless steel lever, the amalgam collection container can
now be removed.
4. Top up the amalgam container to within 5mm of the rim with
neat Puli Jet Plus. Place the container on a solid surface and
carefully place a new lid on top, twist and lock in position.
Then push in the small plastic bung and twist to lock. The
container is now airtight and safe for disposal*. If re-using
the original amalgam container, empty the contents into an
approved amalgam waste container**
5. Turn the drain tap (left hand side of amalgam container) to
the horizontal position.
6. Before replacing or refitting the new container treat the o-ring
seal with a silicon based lubricant (not vaseline).
7. Replace with the new container (043135), place the new
container into position carefully but DO NOT push up onto
the separator.
Relocate the stainless steel lever by pushing downwards to
lock the container into place.
8. Turn the vent valve on the top of the amalgam separator
clock-wise to close it.
1
2
3
7