Model: THD2FE
NON-CONTACT FOREHEAD THERMOMETER
Forehead mode: 93.2 to 108°F/34 to 42.2°C
Surface mode: -7.6 to +176°F/-22 to +80°C
FOR MEASUREMENT OF HUMAN BODY
TEMPERATURE IN PEOPLE OF ALL AGES
Response Time: Forehead mode: 1 second;
Surface mode: 0.6 second
Forehead Measurement Distance: within 1.5 in/4 cm
Surface mode: Distance:Spot = 1:1, emissivity: 0.95
Accuracy: ±0.4°F/0.2°C
Backlit in blue
Memory recalls 25 readings
One-button operation
Battery status indication
Temperature alert or silent
On/Off Power button
Auto-off after 1 minute
ABS plastic
Two 1.5V AAA IEC LR03 alkaline batteries, included
FDA, RoHS, ISO 9001, ISO 13485, ASTM E1965-98,
EN ISO 80601-2-56
Note: Remove label from display before initial use.
Note: In the following instructions, names of the control
buttons are shown in CAPS. Function information that
appears on the display is shown in BOLD CAPS.
USAGE
Indications for Use
The Non-Contact Forehead Thermometer is an
infrared thermometer intended for the intermittent
measurement of human body temperature in people
of all ages.
Intended Operator:
At least 11 years old (5 years intensive reading
experience), no maximum.
Ambient Temperature
If there is any temperature difference between the
places where the device is stored and where it will
be used, allow the thermometer to adjust to the room
temperature where the subject is for at least
15 minutes before measurement.
Points of Attention
1. Forehead temperature is displayed in oral mode.
This mode converts the forehead temperature to
display its “oral-equivalent” value.
2. Before the measurement, the subject should stay
in a stable environment for 5 minutes and avoid
exercise or bath for 30 minutes.
3. Keep the forehead area clean and free from sweat,
cosmetics and scars while taking the temperature.
4. The “Clinical Bias” is -2.5 to -3.1°F (-1.4 to -1.7°C).
5. The “Limits of Agreement” is 0.98.
6. The “Repeatability” is 0.36°F (0.20°C)
BEFORE WE BEGIN
This thermometer has been designed for use in
professional healthcare facility environments and
home healthcare environments (as dened in 3.1 of
IEC 60601-1-11 Edition 2.0 2015-01 and Guidance
for Industry and Food and Drug Administration Staff:
Design Considerations for Devices Intended for
Home Use, Issued November 24, 2014). It is not
meant to replace a visit to the doctor. Compare the
measurement result to your regular
body temperature. Consult with doctor
if you have health concerns.
PO Box 10947 Portland, OR 97296-0947 Tel 800 338-5594 Fax 800 879-2364 Web CDNkitchen.com

Other manuals for CDN THD2FE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CDN THD2FE and is the answer not in the manual?

CDN THD2FE Specifications

General IconGeneral
Accuracy+/- 1°F between 32-212°F, +/- 2°F elsewhere / +/- 0.5°C between 0-100°C, +/- 1°C elsewhere
PowerBattery
Battery TypeAAA (1.5V)
Display TypeLCD
TypeDigital
Resolution0.1°F/°C
DisplayDigital
MaterialStainless Steel
FeaturesAuto Power Off
Response TimeApprox. 5 seconds
Auto Power OffYes