19
8.3 When Troubled
The checkpoint when being troubled. Before judging it as failure
When functional fault arises, check based on the following table.
By the corresponding processing, when fault is not canceled, tell repair to our company or a store.
In addition, I am allowed to make a torque tester's term of a guarantee into one year from a purchase.
(However, there may be onerous in the contents of repair.)
Use a wish attached case and a plastic Case as an object for storage of a measuring instrument.
Moreover, when you convey this product to our company by check, proofreading, repair, etc.,
for product protection, put into this case and is sure to send.
For any inquiry,
Contact your dealer.
-ISO9001-
E-mail sales@cedar.co.jp
Homepage http://www.cedar.co.jp
SUGISAKI METER CO.,LTD.
A power supply is not
turned on.
Carry out regular time charge using
the exclusive charger attached.
Even if it turns on a
power switch, a LOBAT
lamp blinks and a power
supply is shut off
immediately.
Carry out regular time charge using
the exclusive charger attached.
It does not display, even
if it switches on a power
supply after charge of
regulation time.
If it is not used over a long
period of time, an internal
battery will carry out full
electric discharge.
An AC/DC adaptor is connected and a
system reset button is pushed.
After a return, push a clear button and
give an indication zero.
When torque is applied, does a
numerical value change?
The gauge may be damaged when not
changing. Request repair.
Is it the numerical value of
each function under setup?
End a setup and return to measurement
mode.
A display disappears
immediately.(It does not
hold by measurement
mode PP/PD)
It is used setting up an auto
clearance.
Check the setting item of an auto clear
function again. When there is no
necessity, set it as 0.0C.
A zero point cannot be
taken.
Did you clear and make it
zero?
"3.5" is seen and zero adjustment is
performed.
By real-time output
mode, even if it imposes
torque, a display does
not change.
Data are not outputted.
Torque has not reached a
real-time output start value.
At real-time output mode, when
torque is under a "CLO" real-time
output start value, a data output does
not give an indication with "0", either.
Please change the value of "CLO" into
a suitable value.