EasyManua.ls Logo

cefar Direct TENS - Ending Therapy

cefar Direct TENS
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
Direct TENS
4.2. Ending Therapy
The default setting of the Direct TENS™ device
is continuous operation. If you want to end the
therapy, switch o the device with the ON/OFF
switch.
When the therapy timer is activated, the device
switches automatically o at the end of the
programmed interval. The remaining time is
indicated at the top of the display.
Check that the Direct TENS™ stimulator is switched
o before you remove the electrodes.
• If the therapy timer is not activated, switch o the
Direct TENS™ stimulator with the ON/OFF switch.
• Remove the electrodes very carefully. Do not pull
on the leadwires, but on the electrode.
• Reattach the electrode to its protective paper. Be
sure to attach the electrode to the side marked
“on”, not to the side marked “no”.
• Disconnect the electrodes from the leadwire.
• Clean the skin with a mild soap solution.
• Electrodes no longer fit for use can be disposed
of with the normal domestic waste.

Related product manuals