20
This characteristic of the DC112d/DC112 allows measurements to be carried out for workers
with split shifts or special shifts (eg. weekends). It also enables the battery to be changed
without the need to make two measurements.
2.3.6 Ending a measurement
By pressing you end the measurement or recording. The measurement status
indicator will change from
to .
2.3.7 Consulting measured data
While the DC112d/DC112 is not measuring ( ), you may consult all the functions measured.
To consult them, apply the same procedure as described in section
0 on viewing data while
measurement is in progress.
2.4 Overload indicator
The DC112d/DC112 is equipped with an overload indicator for each function. If overloading
occurs during a measurement, the ^ symbol will appear before the function affected by
overloading. When a function registers overload its corresponding measurement may be
incorrect.
During the overload the
indicator appears in the bottom right-hand corner of the screen.
2.5 Menu of the DC112d/DC112: Register management and settings
This section details the options that are accessible from the DC112d/DC112 menu, among
them; memory management, setting the measurement parameters, setting the language, the
time and date and, only in the DC112 selection of frequency weighting for the spectrum
analyser.
The DC112d/DC112 comes with an initial configuration which makes it possible to carry out
measurements without having to reset it in advance.
2.5.1 Menu access
To access the DC112d/DC112 menu, make sure that no measurement is in progress ( )
and press
.
The following screen will appear: