CARE AND MAINTENANCE
FUNCTIONAL CHECK
CHECKING FOR LEAKS
1. Close the valve on the oxygen cylinder.
2. Depressurize the EVOLUTION
TM
by inhaling several times using the nasal cannula. The gauge
indicator should drop to zero.
3. Check that all screwed unions and tube connections are tight. If necessary, tighten them by hand.
NOTE: Do not use a tool.
4. Ensure that the unit is set to the “PULSE” position.
5. Slowly open the valve on the oxygen cylinder until the needle in the gauge indicator no
longer moves.
6. Close the cylinder valve again.
7. Observe the needle in the gauge indicator for approximately one minute.
If the needle remains in its position, everything is OK.
However, if the contents indicator shows a continuous decrease in pressure, there is a leak in
the system. In this event, contact your Home Care Provider.
DISPOSAL
The Unit: Do not dispose of the unit in the household waste. Consult an authorized electronic waste
recycling company for the proper disposal of the unit.
Disposal of Batteries: Do not dispose of used batteries in the household waste. Contact a public waste
disposal authority for proper disposal instructions.