24
CLASSIFICATIONS AND SPECIFICATIONS
SEPARATION DISTANCES
Recommended separation distances between portable and mobile
RF Communications equipment and the EVOLUTION
TM
.
The EVOLUTION
TM
is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or user of the EVOLUTION
TM
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications (transmitters) and the EVOLUTION
TM
as recommended below. According to the
maximum output power of the communication equipment.
Separation distance according to transmission frequency in meters
Max Output Power
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(Watts)
D = (3.5/V
1
)(✓P) D=(3.5/V
1
)(✓P) D=(7/V
1
)(✓P)
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33