10
Power
frequency(50/60
Hz) magnetic
field IEC
61000-4-8
3 A/m 3 A/m
The MEDICAL SCALE
MS4971 power
frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration-electromagnetic immunity
The MEDICAL SCALE MS4971 is intended for use in the electromagnetic
environment specified below.
The customer or the user of the MEDICAL SCALE MS4971 should assure that is
used in such and environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Conducted RF
3 Vrms
Portable and mobile RF
communications equipment
should be used no closer to any
part of the
MEDICAL SCALE
MS4971 including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
27
SCALE ASSEMBLING
Column:
Step 1. Assembling column with base.
Step 2. Connecting wire between
column and base first
Step 3. Tightening four screws at the bottom of base.
Assembling is complete.