Do not exceed the maximum values specified in the applicable
standards.
If you have any problem, contact your local CHARDER MEDICAL service
partner.
ELECTROMAGNETIC COMPATIBILITY
ADVICE TO THE USER
This product is a sensitive electronic instrument and as such it may
temporarily be affected by Radio Transmitting Devices being used in close
proximity to it (such as Mobile Phones, Walkie Talkies, CB Radios, and
Radio Model Controllers). If it does exhibit symptoms such as the display
of erratic or erroneous data, try moving this product away from the source
of interference or switching off that source of interference whilst you are
using this product. If the problem persists then contact your local agent.
WARRANTY-LIABILITY
If a fault or defect is present on receipt of the unit which is within
CHARDER MEDICAL’s scope of responsibility, CHARDER shall
have the right to either repair the fault or supply a replacement unit.
Replaced parts shall be the property of CHARDER. Should the fault
repairs or replacement delivery not be successful, the statutory
provisions shall be valid. The period of warranty shall be two years,
beginning on the date of purchase. Please retain your receipt as
proof of purchase. Should your scale require servicing, please
contact your dealer or CHARDER MEDICAL Customer Service.
No responsibility shall be accepted for damage caused through
any of the following reasons: Unsuitable or improper storage or use,
-
28
-
2. LCD display faulty
Possible hardware defects include: Uneven brightness in the LCD
display screen & texts color blurred, smeared rainbow screen,
incorrect decimal display
LCD PIN broken or short circuit
PCB cooper foil broken & loosed welding
Unable to save or read data – IC or transistor faulty, internal parts
broken.
LCD showing ERRL after switch on - Load cell damaged
Overload may cause the weigh to malfunction.
Software system crash
Resonator faulty
Load cells with faulty grinding standard.
Key buttons failure - Front key panel damaged or disconnected
3. Buzzer malfunction
Wrong welding of PVC wire
Key buttons & control panel damaged or disconnected.