16
Fluidotherapy® - Dry Heat Therapy
SETUP
Press the “INCREASE” or “DECREASE”
buttons until desired Air Speed is
displayed.
NOTE:
The “AIR SPEED” is adjusted in increments
of 5 from 0 to 100.
TREATMENT MODE PARAMETERS (CONTINUED)
TREATMENT TEMPERATURE (CONTINUED)
Press the “INCREASE” or “DECREASE”
buttons until desired Treatment Temperature
is displayed.
NOTE:
°F can be changed to °C and vice versa as
desired. See page 20 for instructions.
Treatment Temperature can be adjusted in 1°
increments from 88 °F to 130 °F
(31.1 °C to 54
°C).
AIR SPEED
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“AIR SPEED”.
NOTE:
The default speed of “50” will be displayed.
PULSE TIME
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“PULSE TIME”.
NOTE:
The “PULSE TIME” allows the unit to
operate by pulsing the medium during
treatment. If selected and set, the medium
will pulse on and off in equal increments.
The Factory Default is “OFF”.
EXAMPLE: “PULSE TIME” is set at “5”.
The unit will pulse the medium during
treatment, five seconds on and five
seconds off.