15
Rated Maximum Output Power of
Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = [3.5]√P
V
1
80 MHz to 800 MHz
d = [3.5]√P
E
1
800 MHz to 2.5 GHz
d = [7]√P
E
1
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
15
SETUP
Use the following instructions to set the various Mode Parameters to
the desired settings.
TREATMENT MODE PARAMETERS
Press the “INCREASE” or “DECREASE”
buttons until desired Treatment Time is
displayed.
NOTE:
The “TREATMENT TIME” can be adjusted in
one minute increments from 1 to 99.
ATTENTION!!!
ATTRITION OF PARTICLES DURING SHIPMENT
MAY RESULT IN INITIAL DUSTING AND SMALL
PARTICLE RELEASE. DUST AND ODOR WILL
DISSIPATE WITH USE.
TREATMENT TIME
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“TREATMENT TIME”.
NOTE:
The default time of “20:00” will be
displayed.
TREATMENT TEMPERATURE
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“TEMPERATURE”.
NOTE:
The default temperature of “115 °F (46.1 °C)
will be displayed when unit is first plugged
into wall outlet.
The “F” or “C” and “TEMPERATURE” indicator
will flash while the programmed Treatment
Temperature is being displayed.
When the Bed Temperature is displayed, the
“F” or “C” and the “TEMPERATURE” indicator
will illuminate steadily.
Fluidotherapy® - Dry Heat Therapy